Research Article - Onkologia i Radioterapia ( 2022) Volume 16, Issue 9

A comparison study of out-of-field photon dosimetry between two Varian linear accelerators

Aswathi Ra1,2, D. Khanna1, Hridya V T, Hridya V T1,2, Sathish Padmanabhan2 and P. Mohandass3
 
1Department of Physics, Karunya Institute of Technology and Sciences, Coimbatore, India
2Department of Oncology, Aster Malabar Institute of Medical Sciences, Calicut, India
3Department of Radiation Oncology, Fortis Cancer Institute, Fortis Hospital, Mohali, Punjab, India
 

Received: 06-Sep-2022, Manuscript No. OAR-22-73872; Accepted: 25-Sep-2022, Pre QC No. OAR-22-73872 (PQ); Editor assigned: 08-Sep-2022, Pre QC No. OAR-22-73872 (PQ); Reviewed: 22-Sep-2022, QC No. OAR-22-73872 (Q); Revised: 24-Sep-2022, Manuscript No. OAR-22-73872 (R); Published: 27-Sep-2022

Abstract

Purpose: The present study compares the components of out of field dose of two linear accelerators and quantifies (i) the variation of out-of-field dose with the detector, (ii) the phantom scatter, collimator scatter and head leakage contribution towards out-of-field dose for two linear accelerators (LINAC), and (iii) the variation of out-of-field dose with Field Size (FS)

Materials and Methods: The out-of-field measurements were obtained from Varian Unique Power Linear Accelerator (VUP) and compared with the Varian Truebeam® STx Linear accelerator (VTB) using PTW chambers of different volumes and slab phantoms. The measurements with different chambers were performed for 10 cm2 ×10 cm2 FS with VUP with collimators 00 and 900 . The out-of-field dose components were measured for a small and intermediate FS for VUP and VTB. The measured results were compared with the TPS calculated.

Conclusion: Higher out-of-field was observed with Semi flex chamber with a collimator 90˚ and with a larger FS. Among the machine, VUP showed a lesser scatter factor. From TPS measurements, it was clear that TPS was not modelled for collimator scatter and head leakage.

Results: Comparing the out-of-field dose contribution semi flex showed a higher dose than other chambers. Compared to the individual scatter component, the phantom scatters component shows high with Semi flex chamber, the collimator scatter with the farmer, and head leakage with Semi flex. With collimator rotation of 900 , Semi flex and pinpoint showed an increase in out-offield dose concerning collimator zero. All the components of out-of-field dose increase with FS. When comparing the scatter components of two Linacs, VUP showed a lesser scatter than VTB.

Keywords

out of field dose, scatter factor, Semi flex chamber, non-target dose

Introduction

The goal of External Beam Radio Therapy (EBRT) treatment is to deliver a conformal and focused radiation beam to a target volume to achieve therapeutic benefit within the Clinical Target Volume (CTV). The uncertainties in positioning require an additional margin for CTV and forming Planning Target Volumes (PTV). The PTV refers to the volume of tissue planned to receive the prescribed dose. The advanced treatment techniques are used to deliver a conformal higher dose to the tumour while minimizing adjacent normal tissue doses with the help of beam shaping. Deterministic effects can be avoided by sparing the normal structures, but it may not reduce the stochastic effects

Radiation has little effect on tissue outside of the PTV. Nontarget doses are split into two categories:

a) A non-target dose that is within a primary field border, such as an entrance and exit dose along the beam path, is referred to as an “in-field non-target dose”

b) “Out-of-field non-target dose” is -a non-target dose that is deposited by stray, or secondary radiation that is not only outside of the PTV but also outside of any primary field edge dose [1].

Any “non-target” radiation should be minimized in radiation therapy as it offers no therapeutic benefit. However, during treatment, unwanted doses are delivered to the non-target volumes in the body. These doses are known as out-of-field doses arising from outside the primary beam. Non-target dose or out-of-field dose can be classified into 3 dose levels. High dose (>50% of the prescribed dose), intermediate dose (5%-50% of the prescribed dose), and low dose (<5% of the prescribed dose). Because of cancer screening and contemporary medicines, the possibility of late consequences from secondary radiation may be more apparent today in the modern day. Many published studies have reported non-target doses from various radiotherapy approaches [1-3]. Dose to the non-target can be reduced in many ways, such as minimizing the size of CTV or PTV by reducing the field size, treatment technique, treatment energy, beam angle, Multi Leaf Collimator (MLC), jaw tracking, and machine shielding. Outfield-field doses arise from several contributions such as head leakage, scattering at the beam collimators, and scattering from the patient body or phantom. Thus the out-of-field dose is the sum of phantom scatter, collimator scatter, and head leakage.

Total out-of-field dose (T)=Head leakage (L)+Collimator scatter dose (S)+Phantom scatter dose (P). ----Equaton 1

Radiation scatters from a patient or phantom are the leading cause of out-of-field dose near the treatment field edges. Phantom scatter depends on the field size and patient characteristics. Leakage radiation has a significant contribution in large distances. MLC may enhance the leakage radiation, and collimator scatter contribution by increasing the Monitoring Unit (MU) and treatment time.

The Treatment Planning System (TPS) is used in EBRT to calculate the treatment dose accurately but does not correctly calculate the dose outside the primary beam due to less optimization. Moreover, the out-of-field dose due to photon, proton, and neutron is a challenging issue. For a 6 MV beam, the average energy outside the treatment field is between 0.2 MeV to 0.6 MeV. Many studies indicated the inaccuracy of outof-field dose calculation algorithms in TPS systems and different cancer patients. Cyriac et al. (2015) studied the accuracy of outof-field dose calculation by measuring the components separately using the Oncentra Planning system [2]. Directing the beam to the phantom gives all the 3 out-of-field components, and directing the beam out of the phantom provides leakage and collimator scatter. Leakage radiation is measured by closing MLC completely. Nevertheless, the obtained results show poor calculation in the treatment planning system

J M Bordy et al. in 2013 also did a similar study of out-of-field dose measurement [3]. Measurements are the same as that of cyriac et al., and they observed that the individual components of out-of-field depend on energy [3].

A.M. Abdelaal et al. 2017 reported a higher out-of-field dose with a pinpoint chamber, and they studied the out-of-field dose with the Source to Surface Distance (SSD), field sizes, energy, and depth [4]. Based on this literature review, there is significantly less or no information on comparing the out-offield dose among machines. Abdelaal et al. in 2020 reported a higher out-of-field dose with the Markus chamber [5]. So this study is to compare all the scatter components of out-of-field dose of two linear accelerators, Varian Unique Power (VUP) and Varian True Beam® STx (VTB), for small and intermediate field sizes. Along with that, our study extended to find a suitable ionization chamber for the measurement of dose outside the beam

Materials and Methods

The measurements were performed using 6MV photon beams on VUP having 120 Millennium (Mi-MLC) and VTB with High Definition MLC (HD-MLC). PTW’s slab phantoms along with farmer-type chamber, Semi flex chamber and pinpoint chamber, and Unidose Electrometer were used for measurements. VUP is the single low photon (6 MV) linear accelerator with 60 pairs of Millennium MLC attached to the gantry head as tertiary collimators with 0.5 cm leaf resolution at the isocentre. VTB is a high-energy linear accelerator that can produce 6 MV, 10 MV, and 15 MV flattened beams, 6 and 10 flattening filter-free photon beams, and four electron energies (6 MeV, 9 MeV, 12 MeV, and 15 MeV). VTB has equipped with an HD-MLC with 60 pairs with 2.5 mm leaf width in the central region for 8.0 cm and 5 mm leaf width in the periphery. Semi-flex ionization chamber 31010 (PTW, Germany) is a vented cylindrical ionization chamber with a volume of 0.125 cc that operates up to ± 400 V. The sensitive volume has a radius of 2.75 mm and a length of 6.5 mm with a total wall area density of 78 mg/cm2 . The central electrode is Al 99.98 with a diameter of 1.1 mm

The farmer ionization chamber is a vented cylindrical ionization chamber with a sensitive volume of 0.6 cm3 and is used for absolute dosimetry. The reference point of the chamber is at 13 mm from the tip of the chamber and is operated up to ± 400 V. It has a dimension of the sensitive volume of a radius of 3.05 cm and length of 23.0 mm, and a total wall density of 56.5 mg/ cm2 . The central electrode is Al 99.98 with a diameter of 1.15 mm. The pinpoint ionization chamber, which is also a vented ionization chamber, is used for small field measurements are operated at a nominal voltage of 300 V. This dimension of the sensitive volume of the radius is 1.45 mm and length of 2.9 mm, and a total wall area density of 84 mg/cm2 . The central electrode is Al 99.98 with a diameter of 0.6 mm.

All measurements were performed in slab phantoms at 5 cm depth with a source-to-surface distance of 95 cm. The dose at 5 cm depth for 100 MU is noted for all chambers for field size 10 cm2 ×10 cm2 . To duplicate the clinical scenario, the field size was defined by MLC and jaws. The dose at a point outside the field is the sum of phantom scatter, collimator scatter, and head leakage. The total dose was measured from 1 cm from the field border to 5 cm, as in Figure 1.

oar-16-9-scatter

Figure 1: Measurement setup for scatter dose

The collimator scatters with head leakage components were measured by keeping the chamber at 5 cm depth and 1 cm from the field border. In the same setup, the collimator leakage was also measured. The phantom scatter, collimator scatters, and collimator/head leakage components were measured on VUP 6MV beams with all three ionization chambers with collimator 0° and 90° to know the variation of these components with chamber and with collimator orientation. All mentioned scattering components are measured for LINAC-2 with collimator zero with Semi flex chamber for 3 cm2 ×3 cm2 and 10 cm2 ×10 cm2

To know the variation of out-of-field dose with chambers, we performed all the measurements with Farmer chamber, Semi flex chamber, and Pinpoint chamber at 5 cm depth with collimator and gantry at 0° and calculated the phantom component, collimator component, and leakage dose for VUP. All three components were measured with a collimator at 0° and 90° and verified, if any, for an intermediate field size of 10 cm2 x 10 cm2 . Also, the same is observed for two different field sizes, 10 cm2 ×10 cm2 and. A small field size of 3 cm2 ×3 cm2 was measured for VTB, and VUP was measured with a semi-flex chamber. The contribution of these concerning isocenter dose was calculated and tabulated for both field sizes. The same measurements were performed in TPS with the scanned images of slab phantoms, and the dose was noted, and the out-of-field dose concerning the isocenter dose was calculated. TPS calculated out of filed dose is then compared with the ionization chamber measurements

Results

The individual scatters components (phantom scatter and collimator scatter and head leakage) of VUP were derived for different detectors and with different collimator orientations (0 Degree and 90 Degree) using equation 1 and then calculated the fractional dose in percentage concerning the isocenter dose and were tabulated in Table 1-3.

Tab. 1. phantom scatter contribution (%) concerning isocenter dose for 6MV beams with different detectors

Chamber Type Collimator angle Distance from field border (cm)
1 2 3 4 5
Farmer 3.464 2.163 1.486 1.083 0.8
90° 3.382 2.079 1.471 1.068 0.8
Semi flex 5.474 2.982 1.855 1.3 0.96
90° 5.559 2.943 1.828 1.318 1.001
Pinpoint 5.01 2.722 1.668 1.181 0.936
90° 5.318 3.006 1.669 1.193 0.972

Tab. 2.Collimator scatter contribution (%) concerning isocenter dose for 6MV beams with different detectors

Chamber Type Collimator angle   Distance from field border(cm)
  1 2 3 4 5
Farmer   2.418 1.169 0.641 0.44 0.355
90°   2.423 1.099 0.581 0.417 0.32
Semi flex   0.646 0.391 0.31 0.237 0.196
90°   0.593 0.365 0.29 0.208 0.145
Pinpoint   0.631 0.38 0.304 0.229 0.181
90°   0.569 0.346 0.277 0.19 0.127

Tab.3. Head leakage contribution (%) concerning isocenter dose for 6MV Beams with different chambers

Chamber Collimator angle Distance from field border(cm)
1 2 3 4 5
Farmer 0.003 0.003 0.003 0.003 0.003
90° 0.02 0.021 0.02 0.019 0.019
Semi flex 0.003 0.003 0.003 0.002 0.003
90° 0.034 0.033 0.032 0.029 0.028
Pinpoint 0.008 0.006 0.006 0.003 0.013
90° 0.028 0.028 0.025 0.025 0.028

The percentage of phantom scatter, collimator scatter, and collimator leakage components for VUP and LINA-2 for 3 cm2 ×3 cm2 and 10 cm2 ×10 cm2 with collimator 0° were tabulated in Table 4 to Table 7.

Tab.4. Out-of-field dose in percentage concerning the isocenter dose for 3 cm2×3cm2 field size for Varian Unique linear accelerator

Scatter component (%) Distance from the field border(cm)
1 2 3 4 5
Phantom scatter 2.045 0.839 0.425 0.269 0.183
Collimator scatter 0.355 0.119 0.046 0.024 0.016
Collimator leakage 0.004 0.004 0.004 0.003 0.003

Tab.5. Out of field dose in percentage concerning the isocenter dose for 10 cm2×10 cm2 field size for Varian Unique linear accelerator

Scatter component (%) Distance from the field border(cm)
1 2 3 4 5
Phantom scatter 5.516 2.963 1.842 1.309 0.981
Collimator scatter 0.62 0.378 0.3 0.222 0.17
Collimator leakage 0.019 0.018 0.017 0.016 0.016

Tab.6. Out of field dose in percentage concerning the isocenter dose for 3 cm2x3 cm2 field size for Varian True beam linear accelerator

Scatter component (%) Distance from the field border(cm)
1 2 3 4 5
 
Phantom scatter 1.968 0.801 0.437 0.272 0.183
Collimator scatter 0.41 0.126 0.042 0.021 0.015
Collimator leakage 0.011 0.011 0.012 0.012 0.011

Tab.7. Out-of-field dose in percentage concerning the isocenter dose for 10 cm2×10 cm2 field size for Varian Truebeam linear accelerator.

Scatter component (%) Distance from the field border(cm)
1 2 3 4 5
Phantom scatter 4.775 2.703 1.776 1.256 0.919
Collimator scatter 1.566 0.809 0.495 0.369 0.293
Collimator  leakage 0.012 0.012 0.012 0.013 0.012
 

The same procedure was performed in TPS with the scanned images of the slab phantom, and a fractional dose of phantom scatter and collimator scatter measured for both Linacs were tabulated in Table 8 and Table 9.

Tab.8. TPS calculated scatter component (%) for Varian true beam

Field size
(cm2)
Distance from the field border(cm)
Scatter component (%) 1 2 3 4 5
3x3 Phantom 1.947 0.73 0.487 0.243 0.243
Collimator 0.608 0.243 0.122 0.122 0
10x10 Phantom 4.696 2.988 2.135 1.601 1.281
Collimator 2.135 1.281 0.854 0.534 0.32

Tab.9. TPS calculated scatter component (%) for Varian Unique

Field size
(cm2)
Distance from the field border(cm)
Scatter component (%) 1 2 3 4 5
3×3 Phantom 1.801 0.84 0.48 0.36 0.24
Collimator 0.48 0.12 0.12 0 0
10×10 Phantom 4.348 2.757 1.909 1.379 1.167
Collimator 1.591 0.954 0.636 0.424 0.212

Discussion

This study reveals that the phantom scatters contribution is high with the Semi flex chamber, collimator scatters with the Farmer chamber, and head leakage with the Pinpoint chamber. If we see the out-of-field dose relative to the central axis dose, near the edges -Semi flex showed the highest dose, whereas if we move away from the edges, all the chambers showed almost the same values. Abdallah et al. found that pinpoint showed a higher outof- field dose when compared to Semi flex, whereas in this study, Semi flex showed a higher out-of-field dose near the edges [6]. The last dose was reported with a pinpoint chamber.

With the change in collimator angle, the out-of-field dose is high with collimator 90Ë? compared to zero. This is following Abdalaal et al., who found that with the Semi flex chamber and with collimator 90, the out-of-field contribution was high, whereas, with the pinpoint chamber, the out-of-field dose was less for collimator 90 [6]. Analyzing individual components, the Phantom scatter shows higher for collimator 90 than zero for all chambers. The collimator scatter component was low with collimator 90 for all chambers, and head leakage increased with collimator 90° for all chambers.

It was also found that the phantom scatters, collimator scatters and collimator leakage contribution varies with field size. As field size increases, all these scatter components increase. When compared with the Varian True beam, Varian Unique showed a lower value of scattering factors. The same was observed with TPS calculated phantom scatter and collimator scatter contribution. The TPS showed zero leakage radiation. The ratio of TPS measured scatter factors with that of ionization chamber measured was calculated and it was found that TPS calculated phantom scatter variation from measured increases with distance from the field border (up to 40% within 5 cm) and a substantial variation was found in collimator scatter and head leakage. Huang et al. studied the out-of-field dose inaccuracies for IMRT in Pinnacle TPS and found a 30% variation near the field edge of 3 cm-4 cm and 100% far from the field edge [8]. They stated that the errors appear to be the underestimation of scattered doses from collimators and leakage radiation. R M Howell et al. measured the out-of-field dose in the range of 3.75 cm-11.25 cm from the field edge and found eclipse TPS underestimated the dose by an average of 40% ± 20% [9].

Conclusion

In this study, it was found that a higher out-of-field dose was observed with the Semi flex chamber compared with other ionization chambers. The influence of collimator rotation was also studied and found a slight increase in out-of-field dose with collimator 90 compared to zero. We found an increase in dose outside the field border with field size. Among the machines,VUP showed a lesser scatter factor compared to VTB. From TPS measurements, it was clear that more than phantom scatter, collimator scatter, and head leakage needs to be modeled separately to improve the accuracy of out-of-field dose in treatment planning systems.

CONFLICT OF INTERESTING

The authors have no conflict of interest to declare.

Funding

This publication was prepared without any external source of funding.

Acknowledgement

We appreciate the help and support of Ms. Athuliya K, Medical Physicist of Aster Malabar Institute of Medical Sciences.

Ethical Permissions

Ethical approval was not necessary for the preparation of this article as this study does not involve any human beings.

References

Awards Nomination

Editors List

  • Prof. Elhadi Miskeen

    Obstetrics and Gynaecology Faculty of Medicine, University of Bisha, Saudi Arabia

  • Ahmed Hussien Alshewered

    University of Basrah College of Medicine, Iraq

  • Sudhakar Tummala

    Department of Electronics and Communication Engineering SRM University – AP, Andhra Pradesh

     

     

     

  • Alphonse Laya

    Supervisor of Biochemistry Lab and PhD. students of Faculty of Science, Department of Chemistry and Department of Chemis

     

  • Fava Maria Giovanna

     

Google Scholar citation report
Citations : 208

Onkologia i Radioterapia received 208 citations as per Google Scholar report

Onkologia i Radioterapia peer review process verified at publons
Indexed In
  • Directory of Open Access Journals
  • Scimago
  • SCOPUS
  • EBSCO A-Z
  • MIAR
  • Euro Pub
  • Google Scholar
  • Medical Project Poland
  • PUBMED
  • Cancer Index
  • Gdansk University of Technology, Ministry Points 20